4.4.3 Miscellaneous Commands
Table 4 lists the miscellaneous commands that control general housekeeping func-
tions of the programmable Hipot Tester.
Table 3
Commands
*IDN?
*OPC
*OPC?
*RST
*TST?
*WAI
:SYSTem: ERRor?
:SYSTem: VERSion?
4.5
Details of Command Reference
This chapter will give a detailed description of each command. The examples of
each command will be provided and the return query will be shown where appli-
cable.
*CLS (no query form)
Function:
Clear all event status data registers. This includes the Output Buffer, Operation
Event Status Register, Questionable Event Status Register, and Standard Event
Status Register.
Syntax:
*CLS
Examples:
*CLS
clears all event registers.
20
Returns instrument identification.
Reports when operation is complete by setting the Operation
Complete bit in SESR.
Reports when operation is complete. Same as *OPC except
returns a 1 to the output queue and dose not set the SESR bit.
Resets the protection levels and states, resets the current
and voltage levels to zero, sets the output off, sets both
Group section and Step position to 0.
Initiates internal self-test and reports results.
Wait to continue. This command forces sequential operation
of commands. This command is required by IEEE-488.1-
1987. The power supply, however, forces sequential operation
of commands by design.
Read the next item from the error/event queue.
Return the SCPI version level.
Hipot Electrical Safety Testers Models H111 and H216
Explanation