Descargar Imprimir esta página

AEMC Instruments HIPOT H111 Manual De Instrucciones página 37

Probadores de seguridad eléctrica

Publicidad

Idiomas disponibles

Idiomas disponibles

:DCWStanding:VOLTage
Function:
Set or query the DCW test voltage.
Syntax:
:DCWStanding:VOLTage <NR2>
:DCWStanding:VOLTage
<NR2>
Please refer to specifications
Return:
<NR2>
Examples:
: DCWStanding:VOLTage
: DCWStanding:VOLTage?
is 1.000kV.
:DCWStanding:CMINimum
Function:
Set or query the DCW test minimum current.
Syntax:
:DCWStanding:CMINimum
:DCWStanding:CMINimum?
<NR2>
Please refer to the specification
Return:
<NR2>
Examples:
:DCWStanding:CMINimum
1.00mA.
:DCWStanding:CMINimum?
setting is 1.00mA.
:DCWstanding:CMAXimum
Function:
Set or query the DCW test maximum current
Hipot Electrical Safety Testers Models H111 and H216
1.000 set the DCW test voltage to 1.000kV.
return 1.000 if the DCW test voltage setting
<NR2>
1.00 set the DCW test minimum current to
return 1.00 if the DCW test minimum current
35

Publicidad

loading

Este manual también es adecuado para:

Hipot h1162129.602129.63