• Press the FIELD button again to enter the next field which is "measurement
upper limit". Use the rotary knob to adjust the desired upper limit of resistance.
The arrow buttons adjust the knob's resolution (0 to 9999MΩ).
• Press the FIELD button again to enter the next field which is "measurement
lower limit". Use the rotary knob to adjust the desired lower limit of resistance.
The arrow buttons adjust the knob's resolution (0 to 9999MΩ).
• Press the FIELD button again to enter the next field which is "test time". Use
the rotary knob to adjust the desired testing time. The arrow buttons adjust the
knob's resolution (0 to 999.9s).
• Press the FIELD button again to return the first field which is "test mode".
• Press the EDIT/SAVE button to save all the parameters.
• When Rmin=0 or Rmax=0, only the measurement value appears without a
limit.
5.8.4 Continuity Check (CNT)
• Press the MENU button to enter the "MENU" mode, then use the rotary knob
and arrow buttons to select a step.
• Press the EDIT/SAVE button to enter the "EDIT" mode. The cursor enters the
"test mode" field. Use the rotary knob to select mode CNT. The output current
is fixed to 0.100A.
• Press the FIELD button to enter the next field which is "measurement upper
limit". Use the rotary knob to adjust the desired upper limit of resistance. The
arrow buttons adjust the knob's resolution (0 to 1.200Ω).
• Press the FIELD button again to enter the next field which is "test time". Use
the rotary knob to adjust the desired testing time. The arrow buttons adjust the
knob's resolution (0 to 999.9s).
• Press FIELD button again to return the first field which is "test mode".
• Press the EDIT/SAVE button to save all the parameters.
54
Functions of the FIELD Button
Test mode
CNT
ê
Measurement upper limit
Rmax=XXXXΩ
ê
Testing time
TEST=X.XXXs
Hipot Electrical Safety Testers Models H111 and H216