Menu Parameter Setup
5.8.1 AC/DC Withstanding Voltage Test (V
• Press the MENU button to enter the "MENU" mode, then use the rotary knob
and arrow buttons to select a step.
• Press the EDIT/SAVE button to enter the "EDIT" mode. The cursor enters the
"test mode" field. Use the rotary knob to select mode V
• Press the FIELD button to edit the next field which is "output voltage". Use the
rotary knob to adjust the desired output voltage. The arrow buttons adjust the
knob's resolution from 0.005, 0.01, 0.1 to 5.000kV for VAC, or 0.100 to 6.000kV
for V
.
DC
• Press the FIELD button again to enter the next field which is "trip current
upper limit". Use the rotary knob to adjust the desired upper limit of leakage
current. The arrow buttons adjust the knob's resolution from 0.01, 0.10, 1.00, 10
to 40mA for VAC, or 0 to 10mA for V
• Press the FIELD button again to enter the next field which is "trip current lower
limit". Use the rotary knob to adjust the desired lower limit of leakage current.
The arrow buttons adjust the knob's resolution from 00.01, 00.10, 1.00 or 10 to
40mA for V
AC
• Press the FIELD button again to enter the next field which is "ramp time". Use
the rotary knob to adjust the desired ramping time. The arrow buttons adjust the
knob's resolution (0 to 999.9s).
52
Functions of the FIELD Button
Trip Current upper limit
Imax=XX.XxmA
Trip Current lower limit
Imin=XX.XxmA
RAMP=X.XXXs
, or 10mA for V
DC
Hipot Electrical Safety Testers Models H111 and H216
Test mode
VAC (VDC)
ê
Output voltage
V=X.XXXkV
ê
ê
ê
Ramp time
ê
Test time
TEST=X.XXXs
.
DC
.
/V
)
AC
DC
or V
.
AC
DC