Ranger™ 2000 Series
Accu
Set the status.
OFF
= disabled
RESULT = the Accumulation result is printed
ALL
= all the Accumulation data is printed
Layout [layOUt]
Set the format of the data output to a printer or computer.
Format
Set the printing format.
MULTI = a multi-line (single column style) printout is generate.
SINGLE = a single line printout is generated
Feed
Set the paper feed.
LINE
= move a paper up one line after printing
4LF
= move a paper up four lines after printing
FORM
= a form feed is appended to the printout
Data Transfer [data.tr]
Output weighing results directly to a PC application.
OFF
= disabled
ON
= enabled
Windows XP Setup:
1. Click Start Menu in Windows XP and click Settings -> open Control Panel.
2. Double click Accessibility Options in Control Panel.
3. Select the General tab.
4. Check Use Serial Keys, and click the Settings button.
5. Select the Serial Port, set the Baud rate to 9600 and click OK.
6. Close the Control Panel.
7. Run Excel to open one blank sheet. Focus the cursor on one item. At this
time, if scale sends data to the PC through the RS232 port, the data will be put
into the cell, and the cursor will automatically move to the next vertical cell.
Notes:
Data Transfer Function is not supported in Windows
software for Windows 7/8 users.
If the weighing value is a negative number, set the target cell in TEXT format.
Otherwise, Excel will not distinguish it as a negative number.
Please do not use this function during continuous printing.
End Print
[End]
Advance to the next menu or return to the top of the current menu.
®
7/8. OHAUS provides SPDC