• Alarm 1
• Alarm 2
• STEL
• TWA or
• Fault.
For each selected channel, the drop-down menu enables voting selection. Voting can be set from 1:1 up to 42:42.
7
Maintenance
• The Control Unit requires no special maintenance or cleaning, apart from verifying that it responds appropriately 
when performing sensor calibration checks.
• For SIL 2 (Safety Integrity Level) sensor applications, the calibration intervals must be reduced appropriately.
• Check the calibration of sensor(s) connected to the Control Unit according to your company's safety manager.
DESCRIPTION
Sensor Extension Board (for channels 5 - 8)
Channel Relay Board
Channel Board 4 - 20 mA
GasGard XL, manual
EMC Filter (to be used with external. 24 VDC supply)
Back-up battery (2.2 Ah Kit)
Back-up battery
Display Board
Lid for housing with touch pad and gasket (w/o screws)
Flat ribbon cable (Main Board to Display Board)
Set of Lid fixing screws
Spare battery for Main Board
Main Board (channel 1 – 4)
Fuse 250 V
Spacers for relay channel board [set of 6]
Screw for Sensor Extension Board (need qty. 13)
Table 1 Accessories and Spare Parts
WARNING!
Electrical Hazards
• Disconnect power to Gasgard XL before openin.
• Maintenance should only be performed by a qualified person
Failure to follow these warnings can result in serious personal injury or death.