Performance Tests and Calibration
5
Adjustment is Recommended
Specifications are only guaranteed within the period stated from the last
adjustment. Keysight recommends that readjustment should be performed during
the calibration process for best performance. This will assure that the U1241B/
U1242B will remain within specification. This criteria for re-adjustment provides
the best long–term stability.
Performance data is measured during Performance Verification Tests and does
not guarantee the multimeter will operate within the test limits unless the
adjustment is performed.
Refer to
Calibration Count
to verify that all adjustments have been performed.
Keysight U1241B and U1242B User's and Service Guide