Performance Tests and Calibration
Calibration Count
The multimeter provides the calibration count information for users to access
through front panel operation. Note that the multimeter was calibrated prior to
shipping out to users. Users are recommended to record the initial value of the
calibration count once receiving the multimeter.
The count value increases by one for each calibration point, from 0000 up to the
maximum of 19999. After the maximum count, the calibration count will be reset
to 0. The calibration count can be read from the front panel after the multimeter
has been unsecured, see the following steps:
1 In adjustment mode, press and hold
2 Take note of the calibration count to keep track of the number of calibration
3 Press and hold
Calibration Errors
The following error codes indicate failures that may occur during a calibration.
The error code is displayed on secondary display.
84
the calibration count viewing mode. The primary display indicates the
calibration count value while the secondary display indicates "Cnt".
counts that has been performed.
mode.
for more than one second to exit the calibration count
Keysight U1241B and U1242B User's and Service Guide
for more than one second to view