Table 5-2
Verification Tests
Step
Test Function
Turn the rotary switch to
1
position
Press
to go to
2
function
3
Press
to go to frequency mode
4
Turn the rotary switch to
Keysight U1241B and U1242B User's and Service Guide
Range
1000 mV
10 V
100 V
1000 V
1000 mV
10 V
V
100 V
1000 V
100 Hz
1000 Hz
10 kHz
position Diode
Performance Tests and Calibration
5520A Output
1000.0 mV
10.000 V
100.00 V
1000.0 V
1000.0 mV, 500 Hz
1000.0 mV, 1 kHz
10.000 V, 500 Hz
10.000 V, 1 kHz
10.000 V, 2 kHz
100.00 V, 500 Hz
100.00 V, 1 kHz
100.00 V, 2 kHz
1000.0 V, 1 kHz
1.000 V, 70 Hz
1.000 V, 1000 Hz
1.000 V, 2 kHz
1.000 V
Error from nominal 1 year
U1241B
U1242B
± 1.4 mV
± 11 mV
± 110 mV
± 2 V
± 10.5 mV
± 20.5 mV
± 105 mV
± 105 mV
± 205 mV
± 1.05 V
± 1.05 V
± 2.05 V
± 10.5 V
± 51 mHz
± 600 mHz
± 3.6 Hz
± 5 mV
5
69