10 CLEANING AND DISINFECTION
Elementary disinfection of the cooling waterlines of instruments
10.1
It is recommended to use Alpron or Sanosil S003 or Dentosept P agent in 1% concentration with distilled water.
The solution is to be poured into the bottle for distilled water and can be used permanently. The agents in 1%
concentration are harmless to the health of the patient. With regular use, the cooling system is maintained in
cleanliness and it is not necessary to use other disinfecting agents. Contact your dealer for details about the
possibilities of the purchase and use of the said products.
Cleaning and disinfection of other parts of the dental unit
10.2
Clean the external surfaces of the unit including the glass surface of the touch screen keyboard with a wet cloth.
Use the INCIDIN FOAM – spray (HENKEL- ECOLAB) agent according to the instructions for use at least once a day or
when the surface of the equipment is incidentally contaminated with biological material.
Caution
When cleaning the floor (PVC floor covering) with a disinfecting agent, it is forbidden to place the foot controller
onto the still wet floor covering. It is forbidden to use agents with detrimental effects on the structure of the
varnish or plastic materials (agents based on phenols and aldehydes).
The manufacturer shall not be taken responsible for damages arisen due to the use of disinfecting and cleaning agents
others than those recommended.
Instruments and handpieces
Cleaning, disinfection and sterilization of the instruments and their handpieces should be carried out according to the
instructions of their manufacturer, furnished with the instrument.