WARNING
NOTE
Water drain
The water is conveyed through the measuring chamber via an open
funnel
and the drain hose installed on the same and into the duct.
Remove the supplied funnel
the recess of the housing designated for that purpose.
Connect the funnel of the Testomat
(internal diameter 12 mm/14 mm)
Lay this hose free of back pressure and without the siphon effect
to the drain
Connect mains and devices
Risk of injury from assembly when voltage present!
Unless you disconnect the power supply before commencing assem-
bly, you risk injury, destruction of the product or damage to system
components.
Disconnect all power to the relevant system component before
assembling the Testomat
When connecting, use only verified cables with sufficient
cross-section.
Disconnecting device for the power supply
The unit has no power switch!
®
Fit the Testomat
EVO TH with a switch as a disconnecting device
for the power supply. Use an appliance switch or a circuit breaker
that meets the requirements of IEC 608947-1 and IEC 60947-3.
The switch must be within easy reach of the user of the Testomat
EVO TH and clearly marked as a disconnecting device for the Tes-
®
tomat
EVO TH.
For disconnection, you can also provide a three-pin plug near the
device, which is clearly marked as a disconnecting device for the
Testomat
®
EVO TH CAL.
Danger of damage due to electromagnetic fields!
If you assemble the Testomat
bles parallel to mains cables or in the vicinity of strong electromag-
netic fields, the device may be damaged or the measurement dis-
rupted.
Keep the connecting cables as short as possible
Lay the connecting cables and mains cables separately.
Connect the device with the protective conductor (at 230/100-240
VAC).
Keep interference voltages away from the Testomat
device – e.g. by using a mains filter.
Shield the device from strong electromagnetic fields.
and accommodate it underneath, in
®
EVO TH with a drain hose
®
EVO TH device.
®
EVO TH device or connecting ca-
Assembly
line
®
®
EVO TH