Error messages/troubleshooting

Error messages/troubleshooting
Errors will be recorded in the error history as far as capacity allows
(after which the oldest message will be deleted).
In addition,
error messages can be stored on the SD
card, if in the
menu
=> Basic program => Function SD-Card => Store
error
has been activated.
Details of errors are always sent via the serial RS32 interface.
Errors are handled differently depending on type:
All errors, for which user access is required to recover the function of
the device, are shown as alerts. In addition, an alarm may be trig-
gered. However, it is not possible to completely disable the message,
since the selection "–" is locked in the menu.
In the event of such error occurring, the device reverts to pause
mode. When these errors are acknowledged with the horn button,
they are deleted and the pause mode can be ended. In the event of
any recurrence, the set action (message or alarm) is again triggered.
For such errors, the time of occurrence is recorded in a log.
All errors that occur only intermittently, which means they impact only
on the analysis process, are cyclically checked by the device and
automatically deleted when no longer present.
If these error messages are acknowledged, only the alarm is deac-
tivated. The error message is not, however, deleted. It is deleted only
when the error no longer occurs after rechecking (namely, during the
following analysis, either automatically or when manually started).
These errors can also be ignored when in the menu
Alarm/message
"-" is selected.
In the Error history the start and end time are logged. Likewise, two
notification messages are written via the serial interface and onto the
SD card with the start and end times of the error (see
Message for-
mat).
Regardless of the case, the analyses are continued.