I. Frequency
Range
20Hz
200Hz
2kHz
20kHz
200kHz
2MHz
20MHz
200MHz
Remarks:
- Input amplitude "a" as follows; (DC electric level is zero)
When 10Hz~40MHz : 200mV ≤a ≤30Vrms;
When > 40MHz : Not specified
J. Temperature (COD. 51254 only)
Degrees Celsius
Range
-40ºC~40ºC
40ºC~400ºC
400ºC~1000ºC
Fahrenheit
Range
-40ºF~32ºF
32ºF~752ºF
752ºF~1832ºF
Remarks:
- Included is a K-Type (nickel chromium ~ nickel silicon) point contact temperature probe
which could only measure temperature below 230ºC. If you want to measure temperature higher
than 230ºC, you must use the rod contact temperature probe.
K. 4~20 mA loop current (COD. 51254 only)
Range
(4~20mA)%
Remarks:
When the readings obtained is:
-< 4mA, the primary display shows LO
- 4mA, the primary display shows 0% ....
20mA, the primary display shows 100%
-> 20mA, the primary display shows HI
62
Resolution
0.001Hz
0.01Hz
0.0001kHz
0.001kHz
0.01kHz
0.0001MHz
0.001MHz
0.01MHz
Resolution
0.1ºC
Resolution
0.1ºF
Resolution
0.01%
Accuracy
±(0.1%+15)
Not Specified
Accuracy
±(3%+30)
±(1%+30)
±2.5%
Accuracy
±(4%+50)
±(1.5%+50)
±3%
Accuracy
±(1%+50)
Overload Protection
1000V
Overload Protection
1000V
Overload Protection
1000V
Overload Protection
0.5A, 250V, fast type
fuse, Ø5×20mm