Zero point adjustment (OFFSET)
12
Quitting the SETUP – Menu
1x
short
For confirming and saving.
13
The instrument returns to the last program used
1x long
The instrument must be adjusted until the nominal value reached!
If „ERR 07" appears in the display the OFFSET value is out of the acceptable
range ( Technical Data). The instrument must be returned to the
manufacturer for checking.
Changing the unit of measurement mm / inch (UNIT)
You can select between „mm" and „inch".
The gage must be switched on. The SETUP – Menu can be activated in each program.
Activating
1x long
„UNIT" will appear in the display
3x
short
Activating UNIT – Menu
1x long
The last used measuring unit will appear in the display / Factory setting: mm
Changing the measuring unit
short
For confirming and saving
1x long
Afterwards you will return to the SETUP - Menu.
Quitting the SETUP – Menu
3x
short
For confirming and saving.
The instrument returns to the last program used
1x long
After changing the unit the tolerance limit must be checked again, as the tolerance limit will
be rounded automatically to the new unit and could be incorrect.
SETUP – Menu
39