Message in the pump display
Scale Overload/Check Scale
Communication Error
Calibration Error
Sensor Error
Motor Error
Low Vac Failed Use Alternative
High Vac Failed Use Alternative
Vac Systems Out Use Alternative
Warning beeps
3 warning beeps
5 warning beeps
5 warning beeps
5 warning beeps
5 warning beeps
3 warning beeps
3 warning beeps
3 warning beeps
Error and Warning Messages
Method
Weight on the container scale exceeds
25 kg (55 lbs) or weight on the bag scale
exceeds 12 kg (26.5 lbs). The weight
must be reduced. System function will
resume when the excess weight has
been removed.
Contact Hologic Technical Support.
Contact Hologic Technical Support.
Contact Hologic Technical Support.
Contact Hologic Technical Support.
An alternative low vacuum source must
be used in order to continue the proce-
dure. Contact Hologic Technical Support.
An alternative high vacuum source must
be used in order to continue the proce-
dure. Contact Hologic Technical Support.
An alternative vacuum source must be
used in order to continue the procedure.
Contact Hologic Technical Support.
en
41