J. Push power switch to the "ON" position.
K. Prior to attaching tubing to GI endoscope, prime the irrigation tubing by pressing the footswitch and make flow
rate adjustments as needed. (Ensure that the end of the tubing is placed over a receptacle to catch the expelled
water).
L. Release the footswitch to stop the water flow.
M. Attach ENDOGATOR™ irrigation tubing and proper adapter to GI endoscope.
N. When procedure is finished, turn the power switch to the "OFF" position. After all procedures are completed for
the day , remove tubing from the pump head and discard disposable water bottle, disposable tubing and
GI endoscope connector.
NOTE:
ENDOGATOR irrigation tubing may be used for 24 hours (discard daily).
8
IFU-EGPMA/REV AD