Installation
Configuring the unit
The configuration is done with VistaConfig,
which is automatically installed during the instal-
lation of DBSWIN or VistaEasy
Start > All Programs > Air Techniques > Vis-
i
taEasy > VistaConfig
EN
US
Click

i
This updates the list of connected devices
If the device isn't found, enter the IP address
i
manually and click Register device manually
Activate the connected device in the Regis-
i
tered column
You can also register multiple devices
Use the ScanX device configuration window to
change the device name, (name), to manually
enter an IP address and to request information
Click

i
28
OBJ_DOKU-0000003578-mehrere-011.indd 28
Entering a fixed IP address (recommended)
To reset the network settings, keep the
unit reset button pressed for 15 - 20 sec-
onds while switching on
Deactivate DHCP
i
Enter the IP address, subnet mask and gate-
i
way
Click on Apply
i
This saves the configuration
Testing the device
You can scan an X-ray image to check if the unit
is properly connected
Select the Test tab
i
Select the unit from the Registered Devices
i
selection list
Select the mode class
i
Select the mode
i
Click on Scan Image
i
For the scanning of an image plate, see "113
i
Scanning the image data via a computer"
8.2 Prepare the X-ray cartridge for
an extraoral image plate
The existing X-ray cartridges used with X-ray
film can be retrofitted to enable the use of image
plates
Remove all intensification plates from the X-
i
ray cartridge
2151100018L40 1903V012
09.04.2019 10:51:15