13 Maintenance

13.1 Recommended maintenance schedule
Do not attempt to carry out any maintenance work.
The internal components of the unit are accessible only after parts are removed with special
tools and may be opened and subjected to maintenance by authorized service technicians of
dealers only
Contact your local authorized Air Techniques dealer for any maintenance Non-compliance
with these instructions can lead to equipment damage and loss of the warranty
Prior to working on the device or in case of danger, disconnect it from the mains (e g pull the
mains plug)
ScanX Intraoral View:
ScanX Classic View:
Maintenance
interval
Annually
Every 2 years
2151100018L40 1903V012
OBJ_DOKU-0000003578-mehrere-011.indd 45
The recommended maintenance intervals are based on using the unit for
50 intraoral images per day and 220 working days per year
The recommended maintenance intervals are based on operation of the
device with 40 intraoral and 10 extraoral images per day and 220 working
days per year
Maintenance work
Visually inspect the unit
i
Check the phosphor storage plates and intraoral/extraoral plate protectors for
i
scratches, and replace if necessary
Check the light protection brushes, cut any protruding hairs and remove
i
them
Check the belt drives, transport belts and springs, and replace them if neces-
i
sary
Disassemble the transport drum
i
Remove dust and dirt from accessible parts
i
Mount the transport drum
i
Carry out a system check
i
Replace the pressure roller unit
i
Replace the transport belts and tension springs
i
Replace the toothed belt
i
Usage