Fault
Error code 10000
Error code 10009
Error code 10017
Error code 2
Error code 78
Firmware not running
Settings (e.g. language)
reset after unit restart
Warning message during
shutdown of the unit
2151100018L40 1903V012
OBJ_DOKU-0000003578-mehrere-011.indd 55
Probable cause
Unit is exposed to too much light
Internal communication error
warning; unit remains ready for
operation
Unit shuts down
System error during startup of the
unit
Memory card full
Fault during memory cleanup
A firmware update has been car-
ried out
Internal communication fault
Faulty configuration file
Not a malfunction
Troubleshooting
Solution
Darken the room
i
Turn the unit such that no light
i
can be directly incident in the
entry slot
Update the firmware
i
Wait until the unit has shut down
i
completely
Switch the unit off and back on
i
again
Update the firmware
i
Transmit image data to the
i
computer
Insert an empty memory card
i
Press and hold the reset button
i
while switching on the unit
Update firmware
i
Press and hold the reset button
i
while switching on the unit
Switch the unit off and on again
i
Switch the unit off and on again
i
Update the firmware
i
Reset the configuration to the
i
factory settings and reconfigure
Update the firmware
i
09.04.2019 10:51:23