Usage
Reading the extraoral image plate
In order not to mix up X-ray images, only
scan the X-ray images from the selected
patient
Remove the plate protector from the X-ray
i
cartridge
EN
Place the plate protector with the phosphor
i
US
storage plate in the entry slot of the transport
drum
CAUTION
Loss of image data caused by stray
ambient light entering the unit
Remove the intraoral/extraoral plate
i
protector only after the phosphor stor-
age plate has dropped into the col-
lecting tray
38
OBJ_DOKU-0000003578-mehrere-011.indd 38
Place a finger on the printed surface of the
i
phosphor storage plate and push downwards
until the phosphor storage plate is drawn in
automatically
Scanning progress is displayed on the touch
screen Image data is saved automatically
The touch screen only shows a pre-
view as an initial impression of the X-
ray image Limitations to image pre-
views occur due to image size and/or
exposure conditions For purposes of
diagnosis, the X-ray image must be
viewed on a diagnostic monitor
After it has been scanned, the phosphor
storage plate is erased and drops into the
collecting tray
Save the X-ray image
i
Remove the PSP and prepare it for recording
i
another X-ray image
Remove the plate protector
i
2151100018L40 1903V012
09.04.2019 10:51:19