Usage
Snap-in the plate guide, matching the size of
i
the phosphor storage plate, using a free feed
slit on the transport drum
EN
US
11.2 X-ray
Intraoral X-ray
The procedure is described using a size
2 phosphor storage plate as an example
Required accessories:
– PSP
– Barrier envelope of the same size as the PSP
– Plate guide matching the size of the phosphor
storage plate
WARNING
Risk of cross contamination when
not using the barrier envelope or
when using the barrier envelope
more than once
Do not use an phosphor storage plate
i
without a barrier envelope
Do not re-use the barrier envelope
i
(disposable item)
Preparing the X-ray
– The PSP has been cleaned
– The PSP is not damaged
– The marker (if present) sticks to the correct
position on the image plate If the marker
peels off, replace the image plate
If using it for the first time or if it has been
i
stored for over a week: erase the PSP (see
"115 Erasing the imaging plate")
34
OBJ_DOKU-0000003578-mehrere-011.indd 34
Completely slide the PSP into the barrier en-
i
velope The black (inactive) side of the PSP
must be visible
O p p o s it e
2
t o w a r d s
Opposite side
towards tube.
2142
2 1 4 2
a
D 4 8 4 7 5 8 7
a
D 4 8 4
Pull off the adhesive strip and close the barrier
i
envelope tightly by pressing together firmly
1
2
O p p o s it e
s id e
to w a rd s tu
b e .
2 1 4 2
a
D 4 8 4
7 5 8 7
Recording the X-ray image
NOTICE
Damage to the phosphor storage
plate (PSP) caused by a sharp-edged
holding system
Only use holding systems that do not
i
damage the barrier envelope or the
PSP
Do not use holding systems with
i
sharp edges
Wear hand protection
2151100018L40 1903V012
2
s id e
t u b e .