WARNING
To reduce the risks associated with
chemical exposure due to pressure barrier
failure, connector failure, purifi cation
media or capsule failure from use of
disinfectants within the product:
• Use up to 1 M Sodium Hydroxide (NaOH)
for pre-use and post-use sanitization and
up to 5% Bleach (NaClO) for post-use
sanitization.
• Do not expose the purifi cation capsule to
base sanitization chemicals for more than
60 minutes.
• Do not pressurize the capsule more
than 5 psi during introduction of
NaOH solution and static soak during
sanitization.
• Wear personal protective equipment
(PPE) to prevent chemical splash.
CAUTION
To reduce the risk of burn or exposure
injuries associated with autoclave
sterilization.
• Wear personal protective equipment
(PPE) to handle hot capsules.
Inlet
Outlet
BV0.3R
- 7 -
Figure 1 - Small
Laboratory Capsules
Flow Direction
Air Vent
Air Vent
Inlet
(Luer Cap)
(Luer Cap)
Outlet
BV1R