• For a two-lead trial, connect the
Medtronic OR Cables into "1-L" (left)
and "2-R" (right). The leads show up
as contacts 1-4 and 9-12 in the Bionic
Navigator 3D software.
Medtronic Snap-Lid Connector Cable
(Model # 355031) or Medtronic Multi-lead
Trialing Cable (Model # 355531).
• If the trial is using the Snap-Lid
Connector Cable or Multi-Lead
Trialing Cable, plug it into the OMG
Connector-M socket labeled "Single
Connector." If 2 four contact-leads are
placed in the Snap-Lid Connector Cable
or Multi-Lead Trialing Cable, the first
and second leads show up as contacts
1-4 and 5-8, respectively, in the Bionic
Navigator 3D software. The 4-contact
leads should be programmed in Manual
Mode.
• If 2 eight-contact leads are used, the
leads show up as contacts 1-8 and 9-16
in the Bionic Navigator 3D software.
4. Place the OMG Connector in a secure location
on a flat surface so that patient movement will
not cause the connector to fall.
5. Align and connect the "1-L" and "2-R" ports on
the Precision Spectra OMG Adapter and the
respective OMG Connector.
6. Making sure that the Precision Spectra ETS is
OFF, connect the Precision Spectra ETS to the
port labeled "CD" on the Precision Spectra OMG
Adapter.
CAUTION
Failure to turn off the Precision Spectra ETS
before connecting the Precision Spectra OMG
Adapter may result in excess stimulation.
Programming the Precision Spectra
ETS
1. Link the Precision Spectra Clinician
Programmer (CP) to the Precision Spectra
ETS per the standard procedure.
2. Use the Bionic Navigator 3D Software to
program the Precision Spectra ETS. Refer
to the list of competitor leads for guidance in
selecting the programming mode to use with
the lead.
Disconnecting the Precision Spectra
OMG Adapter
1. Turn OFF stimulation.
2. Disconnect the Precision Spectra ETS from the
Precision Spectra OMG Adapter.
3. Disconnect the Precision Spectra OMG Adapter
from the OMG Connector.
4. Disconnect the proximal end(s) of the Medtronic/
St. Jude OR Cables from the OMG Connector.
CAUTION
Failure to turn off the Precision Spectra ETS
before disconnecting the Precision Spectra
OMG Adapter may result in excess stimulation.
Cleaning the Precision Spectra OMG
Adapter
The Precision Spectra OMG Adapter can be
cleaned using a mild detergent applied with a
lightly dampened cloth or tissue. Residue from
soapy detergents should be removed with a cloth
lightly dampened with water. Do not use abrasive
cleansers for cleaning.
Note: If no longer needed, the Precision Spectra
OMG Adapter should be returned to Boston
Scientific for disposal.