Interference
immunity test
Magnetic field at
mains frequency
(50/60 Hz) according
to IEC 61000-4-8
Note: U
is the rated voltage before applying the test levels.
T
Table 4 – Electromagnetic interference immunity for non-life-sustaining devices and sys
tems
Directives and manufacturer's declaration – electromagnetic interference immunity
The product 'BionicLink 60X7' is designed for operation in an electromagnetic environment as
specified below. The customer or user of the product 'BionicLink 60X7' must ensure that the
device is operated in such an environment.
Interference
immunity tests
Conducted HF inter
ference according to
IEC 61000-4-6
Radiated HF interfer
ence according to
IEC 61000-4-3
IEC 60601 test level
3 A/m
IEC 60601 test level
3 V
150 kHz to 80
eff
MHz
3 V/m 80 MHz to
2.5 GHz
Compliance level
3 A/m
Compliance level
N/A
12 V/m 80 MHz to
1 GHz
3 V/m 1 GHz to
2.5 GHz
Electromagnetic
environment direct
ive
ruptible power supply
or battery.
Magnetic fields at the
mains frequency
should be equivalent
to the typical levels for
commercial or hospital
environments.
Electromagnetic
environment direct
ive
Portable and mobile
wireless equipment
should not be used at
a lesser distance from
the product 'Bion
icLink 60X7' including
the lines than the
recommended safety
distance calculated
using the equation
applicable for the
transmission fre
quency.
Recommended
safety distance:
d = 0.3 √P for 80 KHz
to 800 MHz
d = 2.3 √P for
800 MHz to 2.5 GHz
Where P is the nomin
al output of the trans
mitter in watts (W)
according to the trans
mitter manufacturer
and d is the recom
Ottobock | 21