KPS-PA70 • Digital clamp meter
• Place transfer switch to required measuring function and range.
• When connecting line, first connect the common test line, then connect char-
ged test line. When removing line, remove charged test line first.

4.10 Current Measurement

Electric shock hazard. Remove the probe from the meter before
measuring with current clamp.
• Measuring switch is placed to position A. At this time, the meter is in AC
current measurement state. Choose appropriate measuring range.
• If you want to measure DC current, press FUNC button to enter direct cu-
rrent measurement state.
• Hold the trigger, open clamp head, clip one lead of measurement circuit to
be tested in the clamp.
• Read the current value on the LCD display..
NOTE:
1) Clamping two or more leads of circuit to be tested simultaneously will not get
the correct measuring results.
2) To get accurate reading, connect the lead to be tested at the center of cur-
rent clamp.
" indicates that maximum input AC current is 1000 A.
3) "
4) In order to improve the measurement precision, in the DC current mea-
surement state, if the LCD display is not zero, press REL back to zero, then
measure.

4.11 Voltage measurement

Electric shock hazard. Pay special attention to avoid shock when mea-
suring high voltage. Don't input voltage more than AC750 RMS.
• Insert black probe to COM jack, insert red probe to INPUT jack, choose
appropriate measuring range.
• Place transfer switch to AC voltage V~ position. At this time, the meter is in
the DC voltage measurement state. To measure AC voltage, press FUNC
WARNING
WARNING
41