Programmer Communication - Boston Scientific Vercise Manual De Uso

Ocultar thumbs Ver también para Vercise:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Programmer Communication

The Clinician Programmer (CP) can communicate with a Stimulator. In order to
begin a programming session, the CP and the Remote Control Infrared (IR) windows
must be aligned.
1. Turn on the CP and the Bionic Navigator™ Software will boot-up.
2. Enter the Clinician Options menu on the Remote Control by pressing the
P
and
buttons simultaneously.
3. Select CP Mode.
4. When the Remote Control displays "CP Ready,"
put the IR Interface and Remote Control into the IR
Interface Holder to keep the IR windows aligned.
Note: Orient the IR Interface with the yellow circle
facing down to maximize communication.
5. Plug the serial end of the USB to Serial Adapter into the
IR Interface.
6. Plug the USB to Serial Adapter into the USB port on
the CP.
7. Position the patient within 60 cm (2 ft) of the Remote
Control to ensure a complete communication link
from the programmer to the Stimulator (see Remote
Control Position on page 47).
Note: A message will be displayed on the Clinician
Programmer if the communication range is not
satisfactory.
For more detailed instructions on how to use the Clinician Programmer and Bionic
Navigator Software with the Vercise™ DBS System, see the DBS Programming
Manual.
Detailed Device Description
Physician Manual
91098825-07 Rev A 67 of 333

Publicidad

Tabla de contenido
loading

Tabla de contenido