4. Install a replacement fuse, Inogen RP#125 (BUSS MDA-12) and reassemble the tip.
Ensure the retainer ring is properly seated and tightened.
Cigarette Adapter Plug
CAUTION For continued protection against risk of fire, only use fuse specified.
Inogen One® G3 Column Change Procedure
NOTE
Column change instructions are only to be used when maintenance is
required and are not intended for practice purpose.
1. Turn off the Inogen One® G3 concentrator by pressing the power button to shut
down the device.
2. Remove the Inogen One® G3 concentrator from the carry bag.
3. Remove the battery from the Inogen One® G3 concentrator.
4. Turn Inogen One® G3 over so that is standing in the
upside down position where the display panel is
facing down and the bottom of the Inogen One® G3
is facing up.
5. There are two columns (metal tubes) on the
Inogen One® G3 concentrator and one is located on
each side of the device.
Retainer Tip
Fuse
Cleaning, Care and Maintenance Chapter 6
Retainer
31