5.2 Permitted processes
Process
Parts
Nasal pillow
Forehead
cushion
Plug
Mask body
Elbow
Headgear clip
Headgear
Mask tube
Tube anchoring
clip
Tube anchoring
strap
1
Compatibility has been proven for the
disinfectants GIGASEPT
and ANIOSYME
®
DD1. Follow the instruc-
tions for use for the disinfectants.
2
Alternative to disinfecting.
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
®
FF, CIDEX
®
5.3 Description of processes
Dishwasher
Hand wash
•
•
•
Disinfect
Sterilize
(alternative to
disinfecting)
Notes on the hygiene treatment and on
the hygiene treatment for a change of pa-
tient can be found in a brochure from our
information portal at
www.loewensteinmedical.de. If you do
not have Internet access, contact
Löwenstein Medical.
OPA
5.4 Following treatment
1. Thoroughly dry all parts following the
hygiene treatment.
2. Repeat the hygiene treatment if
required.
3. Perform a function check.
4. Reassemble mask.
5.5 Change of patient
1. Dismantle mask.
2. Wash mask in hot water using a mild
detergent (e.g. washing-up liquid) so
that no residues remain before disin-
fecting or sterilizing.
3. Subject mask to a hygiene treatment
(see table).
4. Use new headgear.
Cycle at 65 °C. Then
carefully rinse off all parts in
clean water.
Hand wash in warm water.
Use a mild detergent. Then
carefully rinse off all parts
with clean water. Allow to
air-dry.
Clean mechanically in dilute
solution, using a brush, for
1
example. Carefully rinse off
all parts with clean water.
Allow to air-dry.
Steam-sterilize in devices to
EN 285. Temperature
134 °C, minimum dwell time
3 minutes.
15