How to use the Tip Card
) Place the neck of the Tip in the cut out.
2) Check wear of the Tip.
3) See the green, yellow and red line to check wear of the Tip. *See below what each color means. At NSK we recommend
to replace a Tip when the Tip meets the yellow line (wear of mm) to guarantee safe and effective use.
Tip Card
* The Tip Card can be used to check the following
tips : G, G4, G6, G8, G9, P/PD, P0, and P20
CAUTION
Tips are consumables. The efficiency of dental scaling
decreases approximately 25% when the top of the Tip
wears  mm and approximately 50% when it wears 2 mm.
In addition, the vibration condition changes owing to the
wear, which may damage a patient's tooth surface. Check
the Tip wear condition with the Tip Card periodically, and
replace the Tip with a new one in good time.
9. How to Use Tip Cover S (Option)
Grip the Tip Cover S and insert it to the Tip.
To remove, grip the Tip Cover S and the handpiece & pull. (Fig. 25)
The Tip Cover S is not designed for use as a Tip changing tool.
CAUTION
Carefully insert the Tip
into the Tip Cover S. Avoid
injuring the fingers.
Neck
Handpiece
Slit
Green: No wear - Tip is OK
Tip replacement is not necessary.
Yellow: Wear of 1mm - Tip is showing some wear
Tip replacement is recommended.
Red: Wear of 2mm - Tip is badly worn
Tip replacement is necessary.
mm
25%
Decrease
Slit
Tip Cover
3
2mm
50%
Decrease
Efficiency
Fig.24
Tip
Tip-Handpiece Joint
Fig.25