• Surgic XT Plus should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is
necessary, the Control Unit should be observed to verify normal operation in the configuration in which it will be
used.
• When operating this system always consider the safety of the patient.
• Read this Operation Manual before use, and fully understand the functions of each part for operation key.
• Do not attempt to disassemble the Control Unit /Foot Control / Micromotor nor temper with the mechanism.
• Check for vibration, noise and overheating before use and if any abnormalities are found in use, stop using
immediately and contact your dealer.
• Use an electrical outlet that is grounded.
• Do not drop, hit, or subject to excessive shock.
• Do not bend the Irrigation Tube while the water pomp is operating. It could cause tube breakage.
• Do not use bent, damaged or sub-standard burs. The Shank could bend or brake.
• Do not exceed the recommended speed.
• After each Operation, be sure to lubricate and sterilize the handpiece as soon as it's cleaned. Blood coagulation can
cause corrosion and rusting. However, do not lubricate the Micromotor. Oil could generate excessive heat and cause
damage.
• The Control Unit can be cleaned with a moist cloth. Disconnect the power supply before cleaning. The Control Unit
and the Foot Control cannot be sterilized by any method.
• When the product is very frequently used, consider the maintenance of a small
stock of replaceable parts.
• Do not disconnect the Motor Cord from the motor.
• Do not wipe with or, clean or immerse in, high acid water or sterilizing solutions.
• Follow the table below for the motion time of motor and handpiece.
Continuous use over a long time may cause the overheating of the
handpiece, which in turn may result in an accident.
• Grounding reliability can only be achieved when the equipment is connected
to an equipment receptacle marked "Hospital Only" or "Hospital Grade".
• Applied parts for patient and/or operator is Handpiece.
• Turn off the Main Power Switch after each use.
• For service requirements and spare parts contact your dealer.
• The use of NSK genuine pre-sterilized, disposable Irrigation tube Kit is recommended.
Temperature
Use
Between 0-40˚C (32-104˚F)
Store
Between -10-60˚C (14-140˚F)
Transportation
** No moisture condensation in the Control Unit.
** Use at outside of these limits may cause malfunction.
NOTICE
Humidity
Between10-85% RH
Between 10-85%RH
2
Continuous time
ON
OFF
3 min.
10 min.
Power plug below is used in North America area.
Plug Type NEMA 5-15P (Hospital Grade Type)
Atmospheric pressure
Between 700- 1060hPa
Between 500-1060 hPa