Endosee
®
Hysteroscope - User Manual
5.4
Sterile, Single-use Disposable Diagnostic
(Dx) Cannula
The cannula contains a miniature CMOS camera and
light-emitting diode (LED) illumination module at the
tip. The cannula connects to the Handheld Monitor
through an electrical connector for: DC power, image
data transfer and removal and control of the CMOS
camera and illumination module. The cannula has an
inner channel for the infusion of fluid. Fluid infusion is
achieved by attaching either a pre-filled syringe or a
pressurized bag of distending medium to the inflow
port through an extension tube or a tubing set. The
outflow port is at the distal tip of the cannula and
empties around the cannula and out the vaginal vault.
The cannula has a pre-curved tip section at an angle
of 25°± 5°. The cannula tip section should not be
altered. The cannula maximum temperature will not
exceed 41°C. Figure 5 shows the cannula.
1. Handheld Monitor Connector
2. Inflow Port
3. Tip Section
4. CMOS Camera & LED
Illumination Module
Figure 5: Disposable Diagnostic (Dx) Cannula
4
3
2
1
16