Final test values related to the nominal capacity (maximum volume) indi-
cated on the instrument, obtained when instrument and distilled water
are equilibrated at ambient temperature (20 °C/68 °F) and with smooth
operation. According to DIN EN ISO 8655.
Accuracy tolerances for the Transferpette
Volume range
µl
100 - 1000
20 - 200
2 -
20
A* = Accuracy, CV* = Coefficient of Variation
Volume step
A*
≤ ≤ ≤ ≤ ≤ ± %
µl
1000
0.6
500
1.0
100
3.0
200
0.8
100
1.2
20
4.0
20
1.0
10
1.5
2
5.0
®
electronic
CV**
≤ ≤ ≤ ≤ ≤ %
0.2
0.3
0.6
0.2
0.3
0.6
0.4
0.8
2.5
Accuracy Table
20 °C
!
Ex
Increment
Type of tip
µl
µl
1.0
1000
0.2
200
0.02
20
53