Troubleshooting - BSVILLAGE PISCINE Simpool Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 119
Simpool pH model
Programmable digital peristaltic pump
Placing back the peristaltic tube
Step 1 - Position the tube and secure it on the left-hand side
Step 3 - Close the front glass

5.3 TROUBLESHOOTING

Problem: the pump does not turn on, the LED and display remain off
Solution:
3. Make sure the connection to the electrical network has been set up correctly in accordance with the
information provided on the pump's identification plate.
4. The electronic board may be damaged as well: call the Support Service.
Problem: the pump is dosing, but there is no pH variation (OFA alarm)
Solution:
6. Check the product's level.
7. Make sure the pH probe is not damaged.
8. Make sure the foot strainer is not clogged up.
9. Make sure the delivery tube is inserted in the probe-holder correctly.
10. Check the status of the peristaltic tube. If evident malformations are detected, make sure that the material
is compatible with the product being dosed by consulting the chemical compatibility table (see ANNEX D),
and then go ahead with the replacement.
ADSP7000407
Step 2 - Insert it in the bell by turning the roller
Fig. 22- Placing back the peristaltic tube
rev. 1.2
INSTRUCTION AND MAINTENANCE MANUAL
27/03/2015
ENGLISH
63/313

Publicidad

Tabla de contenido
loading

Tabla de contenido