Connection J7 - Accessories And Options Interconnections; Connection J10 - Controller-To-Pump Connection - Varian Turbo-V 1000 ICE Manual De Instrucciones

Ocultar thumbs Ver también para Turbo-V 1000 ICE:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 19
PIN
DESCRIPTION
3-11
START S1 signal 24 V, 60 mA, optically
isolated
output
negative).
The output Voltage will be present when the
START push-button on front panel is pressed
or the remote start is present. or the function
has been requested by RS 232/422/485, until
NORMAL operation is reached.
4-12
START S2 signal 24 V, 60 mA, optically
isolated
output
negative).
The output Voltage will be present when the
START push-button on front panel is pressed
or the remote start is present. or the function
has been requested by RS 232/422/485, until
NORMAL operation is reached.
5-13
OVERTEMP. signal 24 V, 60 mA, optically
isolated output (pin 13 positive pin 5 negative).
The output Voltage will be present when a fault
condition is displayed on the front panel
display, with only in an overtemp condition
6-13
FAULT signal 24 V, 60 mA, optically isolated
output (pin 13 positive pin 6 negative).
The output Voltage will be present when a fault
condition is displayed on the front panel
display,
with
condition
7-14
Analog output voltage (0 - 10 V) of pump
speed (pin 7 positive, pin 14 negative).
8-15
Analog output Voltage of DC current drawn by
the turbopump (pin 8 positive, pin 15 negative).
1 Vdc proportional to 1 A.
9-10
Flow meter alarm, the flow value remained
below the threshold for 10 seconds or more.
∗ Open collector type contacts
Connection J7
Accessories and Options Interconnections
J7 connectors
The output voltage of the connector J7 is 120 Vac and
the maximum current is 1A. Do not connect this plug to
a device with a leakage current greater than 1.75 mA.
(pin
11
positive,
pin
(pin
12
positive,
pin
exception
of
an
overtemp
CAUTION!
The 120 Vac, 1 A maximum (independent of line
Voltage) output Voltage is present when the main
power switch is set to 1 position and after START push-
button is pressed and will remain present until a fault
3
condition is displayed on the front panel display or the
turbopump is stopped.
The connector J7 is for vent device, forepump
connector coils, pump cooling fan etc.
To make connections, remove the plug and wire the
pins (maximum wire size 18 AWG, 1 mm²) as indicated
4
in the figure to obtain the desired capability.
Forepump relay coil is an independent user supplied
item.
Connection J10
Controller-to-Pump Connection
A five-meter long cable is provided to connect the
controller to the pump. The following figures show the
controller output connector configuration where pins:
A-F = upper bearing sensor
B-C-D = 54 Vac 3-phase output to pump motor
stator
E = ground
A-G =lower bearing sensor
Controller-to-pump connector
(applicable to model 969-9464 and 969-9564)
54
TECHNICAL INFORMATION
NOTE
87-900-951-01 (B)