Press the PUMP CURRENT push-button and the
display shows
G A S
T Y P E
S E L E C T I
Select the type of purge gas, the default is Nitrogen. Argon
can be selected if needed.
•
Press the PUMP CURRENT push-button and the
display shows:
H E A T E R
S E L E C T I
Heater Jacket power ON or OFF: setting ON powers on
the Heater Jacket, setting OFF powers it off.
The Heater Jacket automatically switches to OFF when the
controller is in the Failure State.
•
Press the PUMP CURRENT push-button and the
display shows:
H E A T
J A C K E T
S E T T I
N G :
Heater jacket operating temperature selection. The range
is 40-90 °C, set to 65 °C by default.
•
Press the PUMP CURRENT push-button and the
display shows:
L O W E R
T 2
S E N S O R :
Detection of the pump's lower bearing temperature. OFF
will always be indicated if there is no sensor on this
bearing.
and then:
S O F T
S T A R T
S E L E C T I
where: XXX = YES or NO.
If YES is selected, the Soft Start mode allows the pump
to rump-up the Normal speed within ten steps.
When NO is selected, the Soft Start mode is
deselected and the rump-up of the pump will be done
within 360 seconds. The controller is factory set to
YES.
The Soft Start mode may be deselected/selected only
when the pump is stopped.
O N :
N 2
J A C K E T
O N :
O N
T E M P
X X °
C
B E A R I
N G
O F F
M O D E
O N :
X X X
NOTE
•
Press CYCLE NUMBER push-button to select YES or
NO
S O F T
S T A R T
S E L E C T I
•
Press PUMP CURRENT push-button and the display
shows:
W A T E R
If necessary press CYCLE NUMBER to select YES or NO.
The controller is factory set to NO.
63
TECHNICAL INFORMATION
M O D E
O N :
Y E S
C O O L I
N G
N O
87-900-951-01 (B)
?