USE

General

Make all vacuum manifold and electrical connections and
refer to Turbo-V ICE pump instruction manual before to
operating the Turbo-V ICE controller.
!
To avoid injury to personnel and damage to the
equipment, if the pump is laying on a table make sure it
is steady. Never operate the Turbo-V ICE pump if the
pump inlet is not connected to the system or blanked
off.
The input signal P1 connector should be left in position
including the shipping links if no external connections
are made. The forepump and the Turbo-V ICE pump
can be switched on at the same time.
When the Turbo-V 1000 ICE E-series pump is baked
by a membrane pump, the Soft Start mode should be
deselected.

Startup

Plug the controller power cable into a suitable power
source.
The display lights, and shows:
A U T O T E S T
C H E C K
The above message is only displayed when a board
with optional serial ports is installed.
A U T O T E S T
R E A D Y
F O R
S O F T
S T A R T
WARNING!
NOTE
NOTE
NOTE
O K
L O C A L
The controller with the Soft Start mode allows the pump to
ramp-up to Normal Speed slowly with a minimum ramp-up
time of 30 minutes and a maximum of about 55 minutes.
The Soft Start mode is always operative as default mode. If
it is necessary to deselect this mode refer to the following
paragraph.
If the Soft Start mode is deselected, the ramp- up will be
done within 360 seconds and the display changes as
follows:
P U M P
S T A R T
If the soft start mode is deselected, the hard start
routine is activated on the controller. This means that
during pump startup an higher torque is applied (10
times greater than the standard one) only if the pump is
not able to start. The total number of hard starts and
the number of trials at every pump startup for the last 9
starts are stored in the microprocessor.
If the pump is not connected, the display will be as
shown :
C H E C K
After pressing the START push-button, if the P1
connector is not in place with the link or the external
interlock connection are open, the display will be as
shown in the following figure.
P U M P
I
N T E R L O C K
Unplug the controller power cable and verify the P1
connection according to paragraph "Interconnections".
H S
60
TECHNICAL INFORMATION
R E A D Y
B U T T O N
NOTE
C O N N E C T
T O
P U M P
NOTE
W A I
T
I
N G
87-900-951-01 (B)
:
P U S H
H S
I O N
H S