4.2 Function Description
Getting Started – First time use.
For first-time use, follow the steps below:
1. Make sure that your physician or care provider has initialized and configured the appropriate
treatment values into the unit.
2. Make sure that your package contains a face or nasal mask with headgear. If not, your
physician or care provider can recommend the type of mask and headgear you must obtain
3. Attach tubing and mask (see Setup section).
4. Attach power cord to the unit and an electrical outlet (see Setup section). Unit will
power-on automatically.
5. Use the "MENU" button to set the Current Time, Ramp Up Time, and Ramp Starting
Pressure, (see "MENU" Button Section for description of these options).
Note that these values may have been already set by your physician or care provider.
6. Press the "START/STANDBY" button – Airflow to the mask will begin.
7. You can now place the Headgear and Mask over your head and face and begin treatment
4.3 Menu Button - Option Descriptions
The following items describe the options presented to you after you press the "MENU" button on
the control panel.
(1) Ramp Time
Ramp time function allows user to fall into sleep with a lower, comfortable pressure and
helps users gradually get accustomed to increasing treatment pressure. The first selection
of pressing "MENU" is [Ramp XX MIN]. When the "MENU" setting is in [Ramp XX MIN]
mode, press "UP" or "DOWN" button to set the preferred ramp time and press "MENU" for
confirmation. There are 10 adjustable levels, 0, 5, 10, 15, 20, 25, 30, 35, 40 and 45 minutes.
(2) Ramp Starting Pressure
Press "MENU" button to select [Ramp P XX.X] menu, press "UP" or "DOWN" button to
set the preferred ramp starting pressure and press "MENU" for confirmation. The ramp
starting pressure can be changed from 3 cmH
example, if your therapy pressure is 10 cmH
can select is 9 cmH
O.
2
(3) Therapy Pressure
Press "MENU" button to select the [P
pressure setting displayed in cmH
provider, a respiratory therapist or doctor.
NOTE: The therapy pressure is to only be prescribed by a physician.
(4)
Compliance Meter
Press "MENU" button to select the [CM XXXX.X hr] menu. The compliance meter records
the total therapy hours for the device. The compliance meter should be re-set only by the
provider, a respiratory therapist or by a doctor.
O to "Therapy Pressure – 1" cmH
2
O, the maximum ramp starting pressure you
2
XX.XcmH
O] menu. You can view the current
2
O unit. Therapy pressure is adjustable only by the
2
8
FranÇais
Messages d'Erreur/Avertissement affichés à l'écran LCD
Type de message
Erreur:
La fonction
essentielle ne peut
être exécutée.
Avertissement :
NOTE: Lorsque le message d'alerte s'affiche, veuillez contacter votre médecin ou votre
fournisseur de matériel pour télécharger les informations stockées en mémoire et
O. For
2
reparamétrer le compteur.
Définition
Erreur due à des paramétrages
anormaux du système
Erreur due à une panne du générateur des
flux
Erreur due à un paramétrage anormal ou à
un dysfonctionnement de l'horloge
Mémoire en dehors de l'appareil
La mémoire de l'appareil est presque
pleine
Erreur de communication du module de
carte SD.
Extrayez la carte SD pendant le traitement
des données.
La carte SD est pleine.
Inconsistent ID no. of device with SD card
Wrong firmware version of SD card module
SD card is damaged
57
Message à l'écran LCD
Error 001
Error 002
Error 003
Warn 001
Warn 002
Warn 003
Warn 004
Warn 005
Warn 006
Warn 007
Warn 008