English
Appendix A: EMC Information
Guidance and Manufacturer's Declaration- Electromagnetic Emissions:
This device is intended for use in the electromagnetic environment specified below. The user of this device should
make sure it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment-Guidance
RF emissions
Group1
The device uses RF energy only for its internal function. Therefore, its RF
CISPR 11
emissions are very low and are not likely to cause any interference in nearby
electronic equipment.
RF emissions
Class B
The device is suitable for use in all establishments, including domestic
CISPR 11
establishments and those directly connected to the public low-voltage power
supply network.
Harmonic emissions
Class A
IEC61000-3-2
Voltage fluctuations /
Complies
Flicker emissions
IEC61000-3-3
Guidance and Manufacturer's Declaration- Electromagnetic Immunity:
This device is intended for use in the electromagnetic environment specified below. The user of this device should
make sure it is used in such an environment.
Immunity Test
IEC60601 test level
Electrostatic Discharge (ESD)
±6kV contact
IEC61000-4-2
±8kV air
±2kV for power supply
Electrical fast transient/ burst
line
IEC61000-4-4
±1kV for input/out line
±1kV for differential
Surge
mode
IEC61000-4-5
±2kV for common
mode
<5 % U
(>95 % dip in
T
U
)for 0,5 cycle
T
Voltage dips, short
40 % U
(60 % dip in
T
interruptions and voltage
U
)for 5 cycles
T
variations on power supply
70 % U
(30 % dip in
T
input lines
U
)for 25 cycles
T
IEC61000-4-11
<5 % U
(>95 % dip in
T
U
)for 5 sec
T
Power frequency
(50/60Hz) magnetic field
3 A/m
IEC61000-4-8
NOTE: U
is the a.c. mains voltage prior to the application of the test level
T
Electromagnetic
Compliance
Environment-Guidance
Floors should be wood, concrete or
±6kV contact
ceramic tile. If floors are covered
±8kV air
with synthetic material, the relative
humidity should be at least 30 %.
±2kV for power supply
Mains power quality should be that
line
of atypical commercial or hospital
±1kV for input/out line
environment
±1kV for differential
Mains power quality should be that
mode
of atypical commercial or hospital
±2kV for common
environment.
mode
<5 % U
(>95 % dip in
Mains power quality should be that
T
U
) for 0,5 cycle
of atypical commercial or hospital
T
40 % U
(60 % dip in
environment. If the user of this
T
U
) for 5 cycles
device requires continued operation
T
70 % U
(30 % dip in
during power mains interruptions, it
T
U
) for 25 cycles
is recommended that the device be
T
<5 % U
(>95 % dip in
powered from an uninterruptible
T
U
) for 5 sec
power supply or a battery.
T
Power frequency magnetic fields
should be at levels characteristic of
3 A/m
atypical
location
in
a
commercial
or
environment.
18
FranÇais
2. Description du Produit
1. Pièces comprenant :Appareil PPC principal Câble électrique amovible
2. Câble électrique amovible
3. Manuel d'utilisateur
4. Tuyau d'air flexible de 1,8 m de long.
5. Sac de transport
6. Carte SD
REMARQUE : UNIQUEMENT pour le Praticien ou le Technicien pour télécharger les
REMARQUE : Uniquement applicable aux appareils munis d'un lecteur de carte
REMARQUE : Veuillez utiliser la carte SD (de moins de 2 GB) conforme à la norme
Touche DÉMARRAGE/PAUSE
Filtre à Air
typical
hospital
données. Le patient ne doit pas utiliser cette fonction.
SD.
SDHC. Avant de l'utiliser, veuillez la formater sur FAT16 pour garantir
que les données seront bien copiées.
MENU
Écran LCD
SD Card
Prise Électrique
47
FLECHE HAUT
FLECHE BAS
Orifice de
Sortie d'air