2. Remove the plastic block ② from the lamp holder
(or the exhausted lamp in case of replacement)
by loosening the two locking screws ③. (Fig. 38)
3. Insert the mercury bulb ④ (respect the polarity of
the bulb), tighten the locking screws and refit the
lamp holder inside the lamp housing. (Fig. 39)
4. Insert
the
lamp
fluorescence power supply, aligning the notches
on the connectors. (Fig. 40)
5. Insert the power cord into the connector ⑤. (Fig.
41)
Before connecting the power cord, se-
cure the lamp housing cable the the po-
wer supply.
If the power cord is connected before,
there may be a risk of electrical shock.
housing
cable
into
the
Page 27
③
②
③
Fig. 38
④
Fig. 39
Fig. 40
⑤
Fig. 41