Figure 59.- Spectrum Analyser mode functions menu.
When turning the rotary selector clockwise active option moves downwards while
turning it anticlockwise active option moves upwards.
The downward pointing arrow at the bottom left of the menu means that there are
more functions available, to view these turn the rotary selector clockwise. Below we
describe the use of each function and its range of values.

4.10.1.1 Band Switching

Permits to switch from terrestrial (5-862 MHz) to satellite band (900-2150 MHz)
and vice versa.

4.10.1.2 Span

This function enables selecting the displayed screen frequency range in Spectrum
Analyser mode between Full (the entire band), 500 MHz, 200 MHz, 100 MHz, 50 MHz,
32 MHz, 16 MHz, 8 MHz and 4 MHz (the latter one only in terrestrial bands).
To alter the span, select the functions menu, then turn the rotary selector [4] to
the Span function and press it. The screen will show a window with the spans which
can be selected. Turn the rotary selector [4] to the required span and activate it by
pressing the selector again.

4.10.1.3 Reference Level

The reference level corresponds to the level marked by the top horizontal line
appearing on the Spectrum Analyser mode screen. This function enables the reference
level to be defined between 10 and 130 dBµV in 10 dB steps. The default reference
level is 70 dBµV.