2. With the front panel installed on the enclosure and snapped into place, thread
the free end of the wire seal through either the left or right hole in the IND560
front panel, and through the hole in the retaining clip.
3. Thread the end of the wire cable through the hole in the plastic seal (as shown
in Figure 2-61), remove any remaining slack in the wire, and snap the seal
shut.
Sealing wire
threaded through
hole and clip
Sealing wire
threaded through
plastic seal
Figure 2-61: Harsh Enclosure Seal Threaded and Ready to be Closed – Standard
4. Trim off any excess wire.

Internal Sealing of Both Enclosure Types

Both the panel-mount and harsh enclosures are sealed the same way when sealed
internally. Follow the procedures described below.
Main PCB Switch Sealing
1. Ensure that the appropriate approval region has been selected in setup under
Scale, Type, Approval and that the Metrology security switch SW1-1 is in the
"on" position.
Refer to Closing the Harsh Enclosure on page 2-4 for the correct method to use
when replacing the front panel.
2. On the panel-mount terminal, remove AC power and remove the main board to
apply the internal seal. This step is not necessary for the harsh enclosure.
3. Attach the plastic bottom sealing plate to the Main board as shown in Figure
Figure 2-62.
Figure 2-62: Attaching the Plastic Bottom Sealing Plate
Orientation (left) and with Panel Reversed (right)
IND560 Installation Manual
2-37