Pipetting

28
1.
Fitting the tips
Use the correct tips according to the volume
range or the color code. Ensure that the tips
are securely seated.
Pipette tips are disposables items!
2.
Volume setting
a) Push the volume-change protection upward
to disengage (UNLOCK).
b) Select the desired volume by rotating the
volume-setting wheel. Avoid twisting and
abrupt rotating motions during this adjust-
ment.
c) Push the volume-change protection down to
re-engage (LOCK). Note: The volume-change
protection tightens but does not lock volume-
setting wheel.
3.
Align the manifold
The manifold can turn freely in both directions.
4.
Aspirate sample
a) Press pipetting button to the first stop.
b) Hold the pipette vertically and immerse
the tips 2 to 3 mm into the liquid.
c) Let the pipetting button slide back
slowly. In order for the liquid to reach its
end position, leave the tips immersed
for another 1-2 sec.
d) Touch the tips against the container wall.