6.2. BEFORE USE

6.2.1. Visual Inspection before fitting
Check if there is visible external damage on the vacuum regulator and
its associated equipments (including product labels and marking). If it
shows sign of external damage, remove the products from service and
identify its status.
EN
Visually check if the regulator and its associated equipments are con-
taminated; if needed, apply the cleaning procedure according to clean-
ing procedure here under.
Check if the service or the disposal time has not been exceeded, using
GCE or owner's date coding system. If service or disposal time has been
exceeded, remove the product from service & suitably identify its status.
6.2.2. Fitting to medical vacuum pipeline system terminal units
Refer to appendix Nr 1 to get information about quick coupler type on your
product.
Refer to appendix Nr 2 to get information about connecting / disconnect-
ing procedure.
For regulator fitted with a rail clip and a flexible hose, clip the vacuum
regulator on the rail prior to connect the quick coupler.
6.2.3. Fitting of the security jar
Turn the jar top nut clockwise to connect it to the regulator (do not apply
an excessive torque).
6.2.4. Functional test before use and Leakage test
After connection to medical pipeline system terminal unit and its associ-
ated equipment, set the vacuum level control knob in mid position.
Set the vacuum regulator ON by pushing the red button.
Check that there is a significant vacuum level at patient side inlet by
locking the patient side inlet with your finger.
Set the vacuum regulator OFF by pushing the green button and check
if vacuum level is without change at locked patient side inlet by your
finger.
Set vacuum level control knob at minimum vacuum position by turning
it clockwise.
If any leakage is detected, check connections and repeat leakage test.
If any leakage is detected again, apply the "After use" procedure below
and return the valve to GCE for service or repair.
6.3. OPERATIONS
Set the vacuum regulator ON by pushing the red button.
Lock the patient side inlet with your finger.
Set the vacuum level to the needed level for your therapy by turning the
vacuum level in the anti-clockwise direction and checking the vacuum
indicator.
The vacuum regulator is ready to be used.