6.4. AFTER USE
Before disconnecting regulator from the medical pipe line system ter-
minal unit,
Turn it Off, by pushing the green button and set vacuum level control
knob at minimum vacuum position by turning it clockwise
Remove suction tubing connection from patient side
Visually check if the security jar and the microbiologic filter are
contaminated; if so disconnect it and clean/replace according to
section 8.
Refer to appendix Nr 2 to get information about connecting / disconnect-
ing procedure (For regulator fitted with a rail clip and a flexible hose, dis-
connect the quick coupler prior to un-clip the vacuum regulator on the rail).
7. ACCESSORIES
Standard accessories - security jar and microbiological filter.
Instructions for use for accessories are part of the final package.
GCE recommends the use of GCE security jar (pos. 2 in chapter 6.1) and
GCE microbiological filter (pos. 3 in chapter 6.1).
8. CLEANING
8.1. VACUUM REGULATOR
Remove general contamination with a soft cloth damped in oil free oxygen
compatible soap water and clean with deionised water.
Disinfection can be done with an alcohol-based preparation (moistened
wipe). If another cleaning agent is used, check that it is not abrasive and is
compatible with the product materials including labels and gas (see chap-
ter 5.6.) A suitable cleaning agent is e.g. Meliseptol.
Do not immerse the product in any liquid.
Do not expose to high temperature (such as autoclave).
Do not use cleaning solutions contains ammonia.
8.2. ASSOCIATED EQUIPMENTS
Refer to appropriate Instruction for use (collecting jars).
9. LIFETIME, MAINTENANCE AND SERVICE
9.1. LIFE TIME
9.1.1. Serial number and date of production
SERIAL NUMBER COMPOSITION IS THE FOLLOWING: YY MM XXXXX
Y: 2 last numbers of the year
M: month
X: production number
MediEvac+ serial number is placed on plastic body of valve.
9/93