3.3
Menu Functions
Self Check: During Self Check at power-on, the analyzer automatically checks that the program chip, the tray transport
mechanism, the printer connection and optical system are all OK. The tray type is checked (see Sections 4 and 7.1) to en-
sure that it correctly matches the strip type selected in the menu.
Calibration: For requesting calibration with Control-Test M (see Section 5).
Linefeed: Causes paper to advance. The linefeed is stopped by pressing the left function key ("Stop") (see Section 4).
New Series: For starting a new series of measurements at sequence number 1. It is also possible to have an Automatic
Patient ID (see Section 6.1).
Memory: The analyzer can store up to 100 results together with date and time of measurement, sequence number and pa-
tient ID (if entered). Memory is automatically cleared every time the date changes.
After memory has been erased, "NO RESULTS STORED" is displayed when "Memory" is pressed. Pressing the START but-
ton resumes Ready-to-Measure status. "MEMORY FULL" and the options "Print/Send/Clear" are displayed when the mem-
ory is full. Memory must be cleared before the analyzer can resume Ready-to-Measure status.
Print Results: For generating a printout of stored results. The options are as follows:
• "All": All results in memory
• "Last Series": The most recent series of readings
• "Last One": The most recent reading
Printing can be repeated as often as desired. Printing can be interrupted by pressing the left key ("Stop"), for example to
allow a new roll of printer paper to be inserted ("Linefeed"), and subsequently resumed ("Continue"). The analyzer resumes
Ready-to-Measure status when the "Home" key is pressed or when printing has finished.
Send Results: For sending stored results to the serial interface. Options are the same as for "Print Results". Results can be
sent as often as required (see also Sections 6.7 and 9.1).
Clear Memory: Erases results from memory.
Mode: Choice of Print, Measure and Interface modes.
Printer: Printer options are:
• "On": The printer is switched on. Each result is printed once.
• "2 Copies": In Normal mode, each result is printed twice. Note: after serial readings (Fast mode) and when the printout
is a repeat printout (activated by the "Print Results" function), each result is printed only once.
• "Off": For switching off the printer when the printout of results is only required at the end of a series of readings
(activated by the "Print Results" function), or when results are to be sent via the interface to a personal computer or
host only.
Measurement: Choice of measuring mode:
• "NormalMode": Read cycle 70 seconds, incubation within the analyzer (see Section 6.2)
• "FastMode": Read cycle 30 seconds, incubation outside the analyzer (see Section 6.4)
Interface: Choice of "unidirectional", "bidirectional" or "ASTM" data transfer. For further details, see Sections 6.7 and 9.
Logdata: Choice of sending the log files to the host PC or printing out the last 10 logged data. (Only by supervisors in
Authentication mode)
Strip Settings: For selecting test strip type, units and range limit settings.
Strip Type: Choice of test strip type:
• Combur
Test UX ("C-10")
10
• Combur
7
Test ("C-7")
• Combur
5
Test ("C-5")
The Urisys 1100 system leaves our factory configured for Combur
Combur
Test test strips, you will require the appropriate test strip tray (see Sections 4.1 and 11).
5
Gebrauchsanweisung V04 / Operator's Manual V04 / Manual de instrucciones V04
10
Test UX test strips. If you intend to use Combur
7
Test or
35