E18 INVALID OP.ID!
E19 LIST DOWNLOAD FAILED
E20 NO VALID LIST!
CHECK MEASUREMENT
VALUES OBTAINED ARE
IMPLAUSIBLE WHEN
COMPARED WITH THOSE FROM
VISUAL EVALUATION
NO PRINTOUT
THE ANALYZER WILL
NOT READ EVEN THOUGH THE
SEQUENCE NUMBER IS
DISPLAYED
Gebrauchsanweisung V04 / Operator's Manual V04 / Manual de instrucciones V04
Cause: The entered operator ID is not valid.
Action: Enter a valid operator ID.
Cause: The new operator ID list download failed.
Action: No action. After 2 seconds next state starts with old list if there was.
Cause: There is no valid list at all in the device.
Action: Try to download an operator ID list from the host or continue without au-
thenticated operator.
Analyzer prints out software and chip version number and 3-digit numbers with-
out naming the parameters.
Cause: Service function is activated.
Action: Press the "Back" function key to return to the main menu.
Causes: Test strip incorrectly positioned, uncharacteristic test pad colors, proper
incubation intervals not kept to during serial measurements. The wrong test strip,
such as Combur
Test test strips, may have been used. Electromagnetic
10
interference from other devices (see Section 4).
Action: Repeat the measurement with a new instrument-compatible test strip.
Follow the directions carefully and ensure the test strip is correctly inserted (see
Section 6.2). Repeat calibration if necessary. Remove external sources of interfer-
ence, if there are any.
Causes: "Printer: Off" has been selected, or the analyzer is operating in "Fast
Mode" (see Section 6.4), or the printer/software is defective, or the printer is out
of paper.
Action: Insert paper if needed. Choose "Printer: On" to re-activate the printer.
Request a patient report via the "Print" function. If this fails, activate the "Line-
feed" function. If there is still no response, the instrument is defective.
Action: If an AT/PC keyboard is connected, press the Escape key, or switch
the Urisys 1100 system off and back on again.