Error Messages And Troubleshooting - cobas Urisys 1100 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

CAUTION: Ensure that the positioning hole on the side of the tray is absolutely dry (see Figure 17). This
!
hole is used to ensure that the test strip tray is automatically positioned correctly in the instrument.
4. Pick up the cleaned test strip tray and, with the retaining bar nearest to you and closed (down), slide it into the slot be-
low the function keys, so that the near edge of the tray is flush with the near edge of the analyzer.
CAUTION: Be careful not to touch the grey reference pad. Contamination of the reference pad may impair
!
the quality of the results obtained.
5. If you wish to proceed with the next readings directly after cleaning the test strip tray, switch the Urisys 1100 system on
again. During the self check the system will verify that the reference pad is in good condition for reading and that the
positioning hole in the test strip tray (see Figure 17) is free. If not, an error message will be displayed (see Section 8).

8. Error Messages and Troubleshooting

Error messages are shown in the display but are not printed out. Following 5 minutes of inactivity, the analyzer switches to
Standby mode. The error message is displayed again when the START button is pressed. In the event that the instrument is
defective, contact your local Roche Diagnostics representative.
E1 REFERENCE PAD ERROR
MIDDLE!
E15 REFERENCE PAD ERROR
BOTTOM!
E16 REFERENCE PAD ERROR TOP!
E2 WRONG STRIP !
E3 STRIP MEASUREMENT ERROR ! Causes: No test strip is present on the tray, or the strip is incorrectly positioned
E4 CALIBRATION ERROR !
E5 CALIBRATION INVALID !
Gebrauchsanweisung V04 / Operator's Manual V04 / Manual de instrucciones V04
Cause: The middle portion of the reference pad on the tray is soiled or damaged.
Action: Switch off the instrument. Carefully clean and dry the pad. Check if it is
damaged (e.g. scratched, etc.). Insert the tray again, then wait for the self check to
finish. If the error message is repeated, replace the reference pad or use the spare
tray. Recalibrate with Control-Test M (when using Combur
Cause: The bottom portion of the reference pad on the tray is soiled or damaged.
Action: see E1
Cause: The top portion of the reference pad on the tray is soiled or damaged.
Action: see E1
Cause: The test strip used is different from the one for which the analyzer has
been programmed (e.g. Combur
Action: Press the START button. Repeat the measurement with the type of strip
for which the analyzer has been programmed.
on the tray, the urine on the test strip has dried, the test strip has not been
dipped in urine.
Action: Press the START button. Repeat the measurement with a new test strip.
Ensure that all test pads are dipped in the urine sample. Insert the strip correctly
and ensure that the retaining bar is closed properly after START is pressed.
Cause: Calibration values differ from those obtained in the last valid calibration.
Action: Press the START button. Repeat the calibration with a new calibration
strip taken from the Control-Test M container. Ensure that the strip is properly po-
sitioned under the clip on the test strip tray (see Section 5).
Cause: Calibration values are out of tolerance.
Action: Check the reference pad for soiling or damage. Clean if necessary (see
Section 7.1) or use the spare tray. Repeat the calibration with a new Control-Test
M calibration strip (see Section 5). If the error message appears again, the instru-
ment is defective.