CAUTION: To ensure your readings are always accurate, do not set up the Urisys 1100 system in close
proximity to devices that create high-frequency fields, as they may interfere and produce false results.
!
Such devices include, for instance, walkie-talkies, mobile telephones, microwave ovens and diathermic
equipment.
Instrument connection
Power-on
(1) Unpack the Urisys 1100 system and place it on a stable, level surface. Do not expose the analyzer to direct sunlight or
other source of direct light (for example a spot lamp). For measurements of Combur
tray with the type mark C on the reverse side is required.
(2) Connect the AC adapter to the power socket at the rear of the instrument and to a readily accessible AC wall socket.
Insertion of the test strip tray
(3) Pick up the test strip tray, Type "C", and with the retaining bar nearest to you pushed down (see Figure 1), slide
it into the slot below the function keys, so that the near edge of the tray is flush with the near edge of the analyzer
(see Figure 2).
CAUTION: Be careful not to touch the grey reference pad. Contamination of the reference pad may impair
!
the quality of the results obtained.
Use the test strip tray with the mark N on its reverse side if you measure Combur
ally, select the desired test strip type in the menu Setup 1. The Urisys 1100 system checks whether the defined test
strip type matches the selected test strip tray. It may be necessary to configure the system for the desired test
strip type.
Paper Insertion
To release the printer paper cover, press the structured area immediately below the cover (see Figure 4). The cover can
then be lifted back. Place the paper roll in the compartment and pull out the first few centimetres of paper to just beyond
the edge of the compartment. The thermosensitive side of the paper (the outer surface of the paper roll) should be facing
downwards (see Figure 5). Close the cover again by pressing until it locks audibly into position.
ATTENTION:
!
To remove the printed test report, tear off the paper by pulling it in front across the edge.
(4) Switch on the Urisys 1100 system using the on/off switch at the rear of the instrument.
Self Check
(5) The analyzer automatically carries out a self check. The default language setting for the Urisys 1100 system is English.
The instrument is factory set to read Combur
completion of the self check, the tray is transported to the start position and the retaining bar opens (see Figure 3).
"SELF CHECK OK" is printed together with a date and timestamp.
ATTENTION: If the message E9 Wrong Tray! is displayed after the Urisys 1100 system is switched on, press
the START button. The selection menu STRIP TYPE is displayed. If you use the Combur
!
strips, select C-10 by press ing the left function key. If you use Combur
select C-7 or C-5, respectively, by pressing the corresponding function keys.
When the self test is finished a sound is heard from the printer. The display will show Insert Strip. If the paper has already
been inserted, the printout "Self Check OK" is executed with date and time; this happens whenever the instrument is switched
on. If an error occurs, consult the Operator's Manual or contact your local Roche Diagnostics service representative.
Gebrauchsanweisung V04 / Operator's Manual V04 / Manual de instrucciones V04
Test UX test strips. It checks that the correct tray has been inserted. On
10
Test UX test strips, the test strip
10
7
Test or Combur
5
Test test strips. Addition-
Test UX test
10
Test or Combur
Test test strips,
7
5
1
2
4
5
3
39