Fault
The controller cannot be switched on
(no beep, no red or green light signal).
After switching the controller on, no
aerosol is generated, and/or the
controller switches off again after
a few seconds.
The controller switches off even
though there is still medication left
in the medication reservoir.
Info:
Note that about 1 ml of medication
cannot be nebulised and remains
unused in the medication reser-
voir; this residue must be dis-
carded. This is intentional and
does not represent a malfunction.
The device does not switch off
automatically when no more aerosol
is being produced and there is no
more than about 1 ml medication left
in the medication reservoir.
®
eFlow
rapid - 01/17
Possible cause/Remedy
- To switch on, press the ON/OFF button for
approx. 2 seconds.
- Is power plug plugged into mains and the
device (for mains operation)?
- Are batteries inserted correctly?
- Check the charge level of the batteries or
rechargeable batteries!
- Check the connection between the control
unit and the nebuliser.
Has the medication reservoir been filled with
medication?
More than 1 ml is left over in the medication
reservoir:
- Check the charge level of the batteries or
rechargeable batteries!
- The nebuliser handset was not being held
horizontally.
Hold the nebuliser handset horizontally.
- The controller switches off automatically
after 20 minutes.
Press the ON/OFF button again to resume
treatment, so that you can inhale the pre-
scribed quantity of medication.
The treatment session has been completed
successfully and the device can be switched
off by pressing the ON/OFF button.
en
45