Defender 5000 Indicators
Configure Item
Option(Bold is default)
Duration Time
1 ~ 10 s
4.5.2 Average Time Setup
There are two methods to set the averaging time.
The default Average Time is 10 seconds.
Note: When the time is set to 0, the first stable weight over 5d will be displayed.
Averaging time can be set to a value between 0 and 60 seconds.
Method 1
1. Press the button corresponding to the icon
to enter the Dynamic sub-menu for
changing the averaging time.
2. Press the Softkey corresponding to the icon
and then key in the desired value using the
alphanumerical keypad.
3. After setting the averaging value desired,
press the Softkey corresponding to the icon
for confirmation.
4. Press the Softkey corresponding to the icon
to exit the sub-menu.
Description
Note:
1. The load to be accumulated has to be >= 5d.
Another accumulation can't be done until the
pan is cleared (< 5d).
2. Gross weight and net weight can't be
accumulated together when the LFT is ON (no
such limit when the LFT is OFF or the
approved model is OIML). When first
accumulated value is gross weight (net
weight), scale will enter into gross weight (net
weight) accumulate mode. Otherwise, "Gross
and net weight can't be accumulated" will be
displayed.
3. If the LFT is ON, gross weight 0 has to be
reached before a new sample can be
accumulated. If the LFT is OFF, new sample
can be accumulated after reaching gross
weight 0 or net weight 0.
It is the time for the display to remain the dynamic
weighing result after the load is removed.
EN-47